Urology
Volume 64, Issue 2 , Pages 237-240, August 2004

Spiral computed tomography: Role in determination of chemical compositions of pure and mixed urinary stones—an in vitro study

  • Serkan Deveci

      Affiliations

    • Department ofUrology, Başkent University School of Medicine, Ankara, Turkey
  • ,
  • Mehmet CoŞkun

      Affiliations

    • Department ofRadiology, Başkent University School of Medicine, Ankara, Turkey
  • ,
  • M.İlteriŞ Tekin

      Affiliations

    • Department ofUrology, Başkent University School of Medicine, Ankara, Turkey
    • Corresponding Author InformationReprint requests: M. İlteriş Tekin, M.D., Department of Urology, Başkent University School of Medicine, 1. Cadde 5, Sokak No. 48, Bahçelievler, Ankara 06490, Turkey, Turkey
  • ,
  • Levent PeŞkircioglu

      Affiliations

    • Department ofUrology, Başkent University School of Medicine, Ankara, Turkey
  • ,
  • N.Çağla Tarhan

      Affiliations

    • Department ofRadiology, Başkent University School of Medicine, Ankara, Turkey
  • ,
  • Hakan ÖzkardeŞ

      Affiliations

    • Department ofUrology, Başkent University School of Medicine, Ankara, Turkey

Received 18 November 2003; accepted 16 March 2004.

Abstract 

Objectives

To determine the chemical composition of pure and mixed urinary calculi by multislice helical computed tomography (CT) in an in vitro setting. CT is becoming the preferred radiologic examination in diagnosis and management of calculus disease. The management of calculus disease can be facilitated by ascertaining the stone composition. Unnecessary shock wave treatment could be avoided if the fragility of the stones could be predicted at diagnosis.

Methods

A total of 107 stones (86 pure and 21 mixed) were measured in a multislice helical scanner within an air-filled environment. The scans were performed at two energy levels, 80 kV and 120 kV, with a 1-mm slice thickness. The chemical compositions of the urinary stones were assessed on the basis of the differences in the densities measured in Hounsfield units. The stones were assigned to six different groups according to the density measured in Hounsfield units: uric acid, struvite, cystine, calcium phosphate, calcium oxalate monohydrate, and calcium oxalate dihydrate.

Results

The differences in the density measured by multislice helical CT at 120 kV for the six groups of pure stones were statistically significant. The densities of the different regions within the mixed stones were also significantly different statistically from each other. The densities of the different regions measured in the mixed stones were not different from the densities of the corresponding pure stones. This finding supports the idea that different chemical compositions within a stone can be identified by their densities measured by multislice helical CT.

Conclusions

The chemical compositions of both pure and mixed stones can be determined by multislice helical CT in an in vitro setting. The feasibility of in vivo determinations remains to be clarified.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0090-4295(04)00349-8

doi:10.1016/j.urology.2004.03.029

Urology
Volume 64, Issue 2 , Pages 237-240, August 2004